OPEN RING
For annular plication (as is usual in the posterior mitral and anterior and posterior tricuspid annulus), larger amounts of tissue (> 3 mm) are incorporated within each horzontal mattress suture loop underneath the suture-platform

Images
CLOSED RING
For annular plication (as is usual in the posterior mitral and anterior and posterior tricuspid annulus), larger amounts of tissue (> 3 mm) are incorporated within each horzontal mattress suture loop underneath the suture-platform

Images
For annular stabilization without plication(as is usual in the anterior mitral annulus), the sutures should incorporate exactly 3 mm of tissue, corresponding to the precise distance between the 2 holes in each suture-platform

Images
With the Uniring® free of the holder, slide the Uniring® onto the annulus

Images
Once all sutures are placed one-to-one in the Uniring®, cut at the appropriate heat-sealed joints using a sharp blade in the cutting guide on the outside edge of the “first” and “last” suture-platforms

images
Using the suture guides on the holder, place the individual mattress sutures through the suture-platforms leaving the Uniring® on the holder
Use 2/0 sutures (Cardioflon®) for the mitral annulus

images
Place the sizer (with the “M” showing) over the AML with the notches at the commissures With the sizer against the anterior annulus, place the two intercommissural sutures in the annulus corresponding to the desired location of the “first” and “last” suture-platforms as indicated by the half-oval markings on the sizer, depending upon how far around the annulus the surgeon chooses to go

images
Place one suture at each commissure and use lateral traction to straighten the anterior annulus, and place one suture (“middle” suture) in the center of the posterior annulus (P2) Place additional sutures in the annulus if necessary in order to visualize the unfurled anterior mitral leaflet (AML)

images
The number on the right side indicates the total number of suture-platforms needed and therefore the total number of individual mattress sutures required in the annulus Half-oval markings on the sizer represent the position of the suture-platforms on the final dimension of the native annulus
Story
Uniring
Tricuspid
Mitral
Publication
Contact
Home
IMPLANTATION TECHNIQUE

There must be a 3 mm space of tissue between individual mattress sutures to accomodate the flexible joints of Uniring®

Individual braided polyester horizontal mattress sutures are placed one-to-one in the tissue annulus corresponding to the number of suture-platforms needed

The gaps between the “middle” sutures and the “first” and “last” sutures, (and any additional suture used for exposure of the anterior leaflet), should be made up with the exact number of sutures according to the total number of suture-platforms indicated on the sizer